Home
Suction Irrigation Machine
Contents
1. Manufactured and tested according to 230 Vac 120 VA 50 60 Hz BF Ila 5 0 kg 30 x 14 5 x 32 cm 10 40 C 30 80 40 C 70 C 10 90 lt 70 dB A up to 3000 ml min up to l min gt 20 l min z 8 l min 0 5 1 0 bar 0 3 0 7 bar control unit ambient temperature relative humidity ambient temperature relative humidity depending on the instruments and tunbing used depending on the instruments and tubing used short term with 2 Liter secretion bottle permanent overpressure vacuum VDE 0750 1 DIN EN 60601 1 The nameplate contains technical data such as the type and serial number of your unit which must always be specified when ordering replacement parts or requesting other information Date 10 2001 page 23 24 9 TECHNICAL DOCUMENTATION The unit with serial number was released to the customer on Training was conducted by The following person s received training Function Function 9 1 SERVICE TABLE Use additional sheet if necessary Further technical information only upon request for authorized persons Date 10 2001 page 24of 24
2. Suction Irrigation unit PURGATOR 5100 MANUAL TONTARRA TONTARRA MEDIZINTECHNIK Date 10 2001 page 1 of 24 This manual contains proprietary information that is protected by copyright All rights are reserved This manual or excerpts thereof may not be reproduced by photocopy microfilm or other means or otherwise distributed without the express written consent of Tontarra Medizintechnik GmbH Names that are registered trademarks have not been identified as such The absence of such identification should not be regarded as evidence that a name is not registered as a trademark Tontarra Medizintechnik GmbH would appreciate any comments from users regarding possible errors or unclear passages in this manual Since the improvement of our products is an ongoing process we reserve the right to make design changes without notice CE Identification as per Directive 93 42 EEC Your Dealer Date 10 2001 Manufacturer MEDIZINTECHNIK GmbH Kantstr 33 D 78532 Tuttlingen Germany Tel 49 0 74 61 965 76 0 Fax 49 0 74 61 9 65 76 26 e mail info tontarra de Internet www tontarra de page 2of 24 PURPOSE The intended use of this device is to rinse and aspirate cleaning solution during endoscopic surgical procedures CAUTION Federal U S A law restricts this device to sale by or on the order of a physician or other licensed practitioner WARNINGS e This unit may only be o
3. 1 BEFORE you begin operation with cascaded units for the first time use ONLY SYSTEM units At ist front panel the unit is equipped with Luer Lock connectors for the adaption of irrigation and aspiration vessels In order to avoid unintended interchange the tubing set as well as the connectors at the front panel are coded Luer Lock male Aspiration Vacuum outlet Luer Lock female Irrigation Pressure outlet page 8of 24 3 OPERATING THE UNIT 3 1 FRONT VIEW Graphics of application B Luer Lock connector male vacuum Luer Lock connector female overpressure P Power switch Date 10 2001 page 9 24 3 2 REAR VIEW Max AnschluBleistung NetzanschluB VOR ffnen des Gehauses IMMER erst Netzst cker ziehen EXHAUST Label U Exhaust port V Nameplate W Power outlet for connecting other SYSTEM components X Connection to power supply Y Fuses fine wire fuse as per IEC 127 Z DIN 42801 socket for equipotential bonding conductor Date 10 2001 page 10 24 3 3 Operational elements Mains switch Lights up green when the unit is on When the unit has been switched on the pump begins to run immediately if at least one of the two luer lock connections is open This symbol stands for hydrophobic sterile filter that must always be on both luer lock connections in order to protect the patient from infection and the suction unit
4. E Fine wire fuse as per IEC 127 IP41 Mechanical protection protection against granular particles larger than 1 mm in diameter and dripping water entering vertically S Maximum power consumption apparent power Maximum current consumption U Line voltage f Line frequency A Protection Class Type BF Date 10 2001 page 4of 24 Dear customer Please read these instructions carefully before you put your new unit into operation for the first time This will prevent damage that can result from the wrong electrical connection or improper use Use the suction irrigation unit only for the purposes described in these instructions We will assume no liability for damage caused by using the unit for purposes other than those for which it was designed THIS MANUAL DOES NOT CONTAIN DETAILED DESCRIPTION OF ENDOSCOPY AND IS NOT SUITABLE FOR INTRODUCING A BEGINNER TO THIS SURGICAL TECHNIQUE CONTENTS Page 1 PURPOSE 6 2 INITIAL USE 7 3 USING THE UNIT 9 4 FUNCTION CHECK 12 5 APPLICATION NOTES 13 6 CARE MAINTENANCE AND SERVICE 16 7 ACCESSORIES 18 8 TECHNICAL DATA 23 9 TECHNICAL DOCUMENTATION 24 Date 10 2001 page 5of 24 1 PURPOSE According to the regulations of classical surgery this unit is intended to issue rinsing solutions which preferably have been pre heated to body temperature in order to maintain a clear view of the operation field for the operator and also to aspirate secretions and blood or tis
5. about 2 minutes do not stop the pump The level of water in the scaled receptacle should not become altered any more within this timespan e g if there should be a leaking washer of the overflow protection valve the level of water in the receptacle would continue rising Check imperviousness gt Connect the receptacle according to the graphics shown on the front panel of the suction pump gt Close the Luer Lock connector P airtight e g by a wet thumb or a Luer Lock cap gt Close the connector of the receptacle that leads to the patient airtight gt Switch on the pump gt The sounding from the pump has to stop at least after 40 seconds using a airtight 2 liter receptacle and may not be repeated in intervals of less than 30 seconds If there should be any detected or supposed damages to the secretion receptacle it mustn t be used and has to be replaced by a new Date 10 2001 21 of 24 7 3 Disposable Receptacle 5115 The disposable receptacle below can be used as an alternative to the reusable glass or Polycarbonat bottles This eliminates the reconditioning of the reusable bottle and the suction cap preventing the noso cosmic infection risk The hydrophobic filter integrated in the suction receptacle contrary to mechanical overflow valves holds back aerosol and efficiently prevents contamination in the vacuum system of the suction irrigation unit At the same time smoke which
6. contamination The aforementioned valve is a floater and can be found in the guiding cage that is plugged opposite of the Luer connector at the bottom side of the the suction lid The suction lid can be dismantled into 3 parts and this is done by pulling off the cage that guides the floater as shown below Stud for the connection of suction tubing Luer Lock connector VAKUUM for the connection of suction pump Suction lid Internal suction tube foam absorber washer do NOT remove for cleaning purposes overflow protection valve floater guidance for overflow protection valve cage suction container scaled To put the suction lid with integrated overflow protection valve together again just proceed the reversed way Lubrication of the included washers once in a while will improve their lifetime 19 24 www ree 7 21 Cleaning and Disinfection d Cleaning Date 10 2001 Do not use disinfectants solvants or scrubbing products for cleaning the suction container as this may irreversibly damage the components made of plastic e Do not submerge the container in disinfectant solution e Do not use flammable or explosive agents The plastic components of the suction container are made of Polycarbonat that resists humidity and water Despite of this being submerged in water at a temperature that exceeds 60 C this may result in chemical destruction which af
7. from pollution This symbol stands for the original glass suction receptacle Other vacuum resistant suction containers that are suitable for medicine technical disciplines may also be used The symbol stands for suction rinsing instruments common on the market that have self neutralising trumpet valve to contol the liquid flow The symbol stands for a sterile rinsing solution in a 1 litre glass bottle with fitted puncture tube Puncture tubes for other bottle sizes are available upon request Date 10 2001 page 11 of 24 4 FUNCTION CHECK The function check helps to ensure that the unit is ready for use This should be done prior to any clinical application Control Unit gt Connect unit to power supply and switch on power switch e Power switch lights up in green e When even one of the Luer Lock connectors is left open the pump is working humming sound gt Close both luer lock connections so that they are airtight e g with moistened finger e Pumping noise immediately stops pump becomes turned off e Free the connections alternately The pump begins to operate again Pump noise can be heared Suction lid gt the suction cap without connected tubing set and secretion collection bottle 180 i e upside down e The suction lid must be dry e If there is an integrated mechanical valve e g a swimmer this must be easily movable and must close itself due to the force of gravity gt Clos
8. instruments respectively the instrument combination depending on the intended use e g stronger or softer spray etc Date 10 2001 page 14 24 The suction tube used must at least a bypass opening that prevent tissue being sucked up If only either suctioning rinsing is required the function that is not needed be Switched off by permanently closing the unused Luer Lock connector e g with a Luer Lock cap The pressure or vacuum control will then only occur at the activated connection NOTES Date 10 2001 page 15 24 6 CARE MAINTENANCE AND SERVICE Care Cleaning In addition to proper operation and servicing effective protection of the unit against damage includes setting the unit up safely at the site This means securely setting up the unit on its supporting surface but also includes protecting the unit from moisture dirt and contact with flammable or explosive materials To ensure that the heat generated during use dissipates properly do not cover the ventilation slits The exterior surface may be treated with a cleaning agent or disinfectant that does not damage the paint Avoid using flammable or explosive agents If you must use such an agent wait until it has completely evaporated before switching the unit on again Make sure that no cleaning agents ever get inside the unit Sterilizat
9. is produced by the use of laser is also held back This disposable receptacle 5115 is made of environmentally neutral plastics PP PE and can be burnt when all openings B C D E are closed with the corresponding F caps The negative pressure tube that leads to the P connection on the PURGATOR is connected at the joint piece A This joint piece is pushed onto the connection B which is marked ONLY VACUUM The suction tube depending on the cross section can be connected to one of the openings C D or E The openings that are not used should then be closed with the caps F available After the suctioning process is finished the connection piece B should also be closed with a cap after the joint piece has been removed The suction receptacle volume 2 liter has a scale of 50 ml intervals and the exact amount of secretion quantity that has been removed can be established Date 10 2001 page 22 of 24 8 TECHNICAL DATA Line voltage Maximum power consumption Frequency Protection class Protection symbol Classification as per Annex IX of Directive 93 42 EEC Weight Dimensions H x W x D Operating environment Storage and transportation environment Noise emission Rinsing performance Aspiration performance liquids e g NaCl solution Aspiration performance gas d Pressure in Rinsing bottle Vacuum in Secretion receptacle
10. e operating safety of the unit If desired we will provide additional technical descriptions on request or where we feel it is required for service and repair work Enter all service work performed in the table Technical Service Information Repairing and Returning the Unit Warranty If repairs are required please inform us or your authorized dealer Pack the cleaned uncontaminated unit use only the original packaging and return it to us postpaid Describe the fault or the malfunction and name the person who we may contact in the event of any questions For the protection of the service personnel and for safety during transportation all devices and accessories that are sent in to be repaired must be prepaired for shipment as described in the manual If this is not possible the product must be clearly marked with a contamination warning and should be double sealed in safety foil The manufacturer has the right to refuse to carry out repairs if the product is contaminated Our products are warranted to be free from defects in workmanship and materials for 3 years from date of sale Costs and risks of transportation are not included in this warranty The manufacturer declines all liability for the operational liability and safety of the product if it is opened repaired modified or in any way tampered with by an unauthorized person During the warranty period such actions would render any warranty claim null and void W
11. e the secretion receptacle with the suction lid e Make sure that the suction lid fits tightly e The suction lid should fit into the bottle neck smoothly maybe the O rings on the suction lid need to be lubricated with silicon oil if so take care of sterile conditions Complete System gt Build up the complete system according to the graphic on the front of the unit don t forget the filters and turn on the unit e The pressure ratings must be reached within 45 seconds and the pump is turned off if not check for leakage e The irrigation tubings must be flushed by the instigation of a rinsing process outside the body e Check the suction process in the atmosphere sucking noise pump begins again Do not use the unit if faults or malfunctions are present or d suspected or the unit is obviously damaged Have the unit repaired by authorized service personnel Date 10 2001 page 12 24 5 APPLICATION NOTES 5 1 Functional principles Rinsing The filter frees the air from bacteria and this is pressed onto the surface of a sterile rinsing liquid which is a pressure resistant container bottle and thus this solution is forced out of the rinsing solution container in the direction of the patient H the rinsing solution container is empty the unit carries sterile air approx 4 Itr min to the patient that is if the sterile filter is fitted correctly This means that if an unsuitable or faul
12. fects stability and reduces shock resistance capability of the product gt Dismantle the main components of the secretion receptacle as shown in the sketch gt Clean the surfaces of the components with warm and sterile water from coarse contaminants maximum temperature 60 C and then dry carefully with a soft and smooth clothing gt Maximum cycle time for sterilization steam or dry is 15 minutes at a temperature of 121 C at 1 bar overpressure 100 kPa Please ensure that the opening of the container is positioned downwards During sterilization the components mustn t be squeezed by kind of loads or weights gt After sterilization in an autoclave the components should be allowed to receive ambient room temperature before being used again gt Before reassembling visually check the components on obvious damages gt The receptacle is now prepared f r being used anew The suction container may be sterilized according to the said procedures Maximum is 20 cycles page 20 of 24 7 22 Functional check The suction receptacle is to be checked at least every 6 months or according to hosptal guidelines by trained personnel Overflow valve gt Connect the receptacle according to the graphic shown on the front panel of the pump gt Suc distilled water from a vessel using the tubing that leads to the patient until the overflow protection valve becomes activated gt Continue suction for
13. hen it reaches the end of its service life this unit should be disposed of according to the appropriate regulations Date 10 2001 17of 24 7 ACCESSORIES 7 1 Overview NO Description SI 5100 Suction Irrigation unit Operating Manual Power cord 2 5 m angled as per CEE7 Standard Sheet VII Bottle holder for secretion receptacle 2 liters Polycarbonat Bottle holder for for 1 liter rinsing liquid bottle Suction receptacle 2 liters made of Polycarbonat Hydrophobic sterile filter with LL connectors at both sides 2 pcs Puncture tube Set of silicon tubing 4 pcs 1x 0 8m LL female 1x 0 8m LL male 2 x 2 5m Optional Equipment Replacement Parts NO Description Service Manual Power cord 0 25m straight acc to EN 60320 Luer Lock connector male for tubing ID 5mm Luer Lock connector female for tubing 10 5mm Disposable receptacle 2 L type HI TEC Bottle holder for disposable receptacle Bottle holder for secretion bottle 21 5114 Secretion bottle 21 Schott made of glass Suction lid complete for 5114 Hydrophobic sterile filter with LL connectors at both sides Date 10 2001 page 18 24 7 2 Suction receptacle ACC3011 13 The integrated valve in the suction lid is intended as protection for the pump so that secretion is not sucked into the apparatus causing
14. ion Date 10 2001 CAUTION The following procedures must be performed after each use of the unit and its accessories Disposable filters tube sets and or instruments are designed for one time use and must be disposed of according to the appropriate regulations Silicone tubes Luer lock adapters the supplied receptacle made of glass the suction lid dismantled the puncture tube and metal instruments are sterilizable after being cleaned of coarse contaminants and then rinsed in clean water and carefully dried inside and outside For the treatment of the polycarbonat receptacle refer to item 7 21 They can be sterilized in an autoclave dry air at 134 by steam sterilization at 134 See the respective manufacturer s operating instructions for details Silicone tubings should only be sterilized as often as labeled on its package Never exceed the number of uses recommended by the manufacturer page 16of 24 Inspection The unit must be inspected by an authorized service technician at least once a year In accordance with the special safety regulations for medical appliances all service work such as annual inspection repair alterations calibration etc may be performed only by the manufacturer or persons expressly authorized by the manufacturer If the user opens and endeavours to repair the unit or alterations are carried out by unauthorized persons this shall release us from any liability for th
15. perated by properly qualified persons who have been trained in its use Endoscopic instruments and equipment may only be used by physicians who have completed appropriate training e Explosion Hazard Do not use this unit in the presence of flammable gases i e anesthetics e Electrical Shock Hazard Refer servicing and calibration to qualified service personnel e When this device is used along with High Frequency units only non conductive rinsing solution electrolyt free is permitted to be used Due different vicosity the performance of suction of gas differs strongly from the performance of suction of liquids e When this unit is used 1 6 it is strongly recommended to use an insufflator with audible and optical overpressure alarm capability e Only fully leak proof instruments with self neutralising valves may be used in combination with this suction irrigation unit CONTRAINDICATIONS This device must not either be used for intrauterine distension nor for the distension of other cavities in human beings The use of this device is contraindicated whenever forced suction and irrigation is contraindicated Date 10 2001 page 3of 24 ICONS GRAPHICS AND SYMBOLS d Please note l Caution See accompanying documentation Hydrophobic sterile filter P Connector for the adaption of rinsing equipment overpressure P Connector for the adaption of aspiration equipment vacuum Nameplate SG
16. rself familiar with this system ahead of any initial operation Perforation with the puncture tube should not be done when the rinsing liquid bottle is in the bottle holder but the bottle should be placed on a stable surface Easy replacement of an empty bottle Simply remove the puncture tube from the empty bottle without disconnecting the adapter tubings and puncture a new one d TAKE CARE OF STERILE CONDITIONS e Always begin each OP with a fresh and for this full and sterile solution bottle e Always have a reserve bottle with rinsing solution preferably pre warmed to body temperature ready near the unit When the suction rinsing apparatus has been turned off at the mains the under pressure conditions remain in the connected bottles receptacles Therefore always remove the tubes that lead towards the apparatus first so that the connected bottles can release pressure The same applies when bottles are exchanged during an operation e The adaption of hydrophobic sterile filters at the inlet P and outlet P of the apparatus is obligate to avoid contamination of saline solution and of the vacuum system of the apparatus e Rinsing pressure and suction vacuum are always permanently existent when the unit is switched on This means that only suction and aspiration instruments with integrated i e self neutralising valves can be used The surgeon must choose the type and model of suction and aspiration
17. side of the unit case with the wheel screws included If the pump is fitted onto an infusion stand the bottle holder should not be screwed onto the unit for stability reason By means of a optional fitting kit the bottle holder can alternatively be fixed directly onto the groove of various common unit trolleys The latter method has the advantage that the containers are more easily accessible if the suction rinsing pump is integrated into a tower system page 7 of 24 Connection to Power Supply Connectors Date 10 2001 BEFORE beginning operation for the first time check whether the site voltage matches the voltage range specified on the nameplate The power connection must have a grounding contact Please use the power cord supplied with the unit or an equivalent cord The insufflator is equipped with a rear panel socket that complies with DIN 42801 Use this socket to ground the unit according to local safety regulations Units of the SYSTEM line can be cascaded from the power supply e each additional unit may be connected to the previous one This reduces the number of cords required and allows you to switch all connected units on or off at the same time The TOTAL connected loads of all cascaded units see nameplates may not exceed the connected load specified on the rear panel of the unit Authorized service personal must verify that the cascade complies with electrical threshold values specified by IEC 601 1
18. sue particles with the contaminated rinsing solutions This manual describes an irrigation aspiration pump that gives the user various methods of use but is simple to handle d This device must not be used for intrauterine distension nor for the distension of other cavities in human beings Hint The unit is tested according to and complies with DIN EN 60601 1 2 Nevertheless we recommend to turn off unused equipment around Date 10 2001 page 6of 24 2 INITIAL USE Receiving Inspection Setting up the Unit Date 10 2001 Inspect the unit and accessories included for obvious damage or missing parts immediately after receiving them Damage claims can only be acknowledged if the supplier is notified immediately within 24 hours Always use the original packaging if you must return the unit or accessories Describe the fault or malfunction and name a person who we may contact in the event of any questions Place the unit on a clean and level surface Ambient temperature should be between 10 C and 40 C and relative humidity between 30 and 85 The suction irrigation unit can also be attached to an approved stand such as Item 0500 instead of setting it up on a tower or carrier cart The control unit is fastened to the bearing plate 0510 with two M 4x16 milled screws with 12 to 16 mm of thread To ensure stability all components should be fastened below the sleeve The bottle holders can be fastened to the
19. ty Instrument is used e g if the closing mechanism is not closed properly or leaks and even if the rinsing solution is used up this can result in an unwanted insufflation source which may not have been noticeable at first d Only suction and rinsing tubes with self neutralizing trumpet valves may be used Suction An under pressure secretion reception container is partly evacuated by a pump If the suction conduit is open e g by means of a trumpet valve the under pressure maintains a suction process in the direction of the container Because of the different densities of the suction material gas liquids the suction capacity varies considerably In order to avoid an uncon trolled collapse of the pneumoperitoneum gas should only be sucked off periodically Please take care to notice the pressure indication of the insufflator used during the operation by endoscopic or laparoscopic surgical operations Date 10 2001 page 13 24 5 2 How to proceed Do not use the unit if possible contamination is suspected faults or malfunctions are present or suspected or the unit is obviously damaged Have the unit repaired by authorized service personnel The procedure for conventional open surgical operations and endoscopic surgical operations using suction rinsing units is documented in the appropriate literature This section is only intended to draw your attention to some unit specific procedures D Make you